ReqVise

View TDS

General Information

TDS Code
TDS-FH-001
Title
Fume Hood 4ft Standard Specifications
Category
Fume Hoods › Standard
Project
Created By
AP
Ananya Patel
Created Date
Jan 20, 2025
Revision
Rev 2
Last Modified
Jan 25, 2025
Status
Approved

Description & Application

Description
Technical data sheet for 4-foot standard bench-mounted fume hood with vertical rising sash. Designed for general chemistry laboratory applications with ducted exhaust. The unit features a completely seamless interior liner made of chemical-resistant epoxy resin, providing superior corrosion resistance against acids, bases, and organic solvents at ambient temperature.
Application
Suitable for general chemistry, organic chemistry, and analytical chemistry laboratories. Ideal for institutions, pharmaceutical R&D centers, and quality control laboratories. Designed to handle standard chemical procedures involving volatile compounds, with provisions for electrical outlets, gas, and water supply inside the work chamber.

Applicable Standards

3 standards
IS 4209:2014 EN 14175-3:2019 ASHRAE 110-2016
Tested and certified per IS 4209:2014 for containment performance. EN 14175-3 type test conducted at NABL-accredited laboratory. ASHRAE 110 tracer gas test results available in test report attachment.

Dimensions & Material

Overall Dimensions (L x W x H)
1220 x 900 x 2400 mm
Work Surface Height
900 mm
Sash Opening
500 mm
Weight
280 kg
Primary Material
SS 304
Finish
Powder Coated RAL 7035

Performance Specifications

Parameter Value Unit Test Method
Air Face Velocity 0.5 ± 0.05 m/s ASHRAE 110
Containment < 0.05 ppm SF6 EN 14175-3
Noise Level < 55 dB(A) IS 4209
Power Rating 2.2 kW, 230V / 50Hz IS 4209
Illumination > 500 lux EN 14175-3

Quality & Inspection

Inspection Criteria
Visual inspection of all welds and joints for defects
Dimensional verification against drawing tolerances (±2mm)
Airflow velocity measurement at 9 points on sash opening
Sash operation smoothness and counterbalance test
Powder coating adhesion test (cross-cut) per IS 101
Acceptance Criteria
Face velocity uniformity within ±10% of average at 500mm sash height
Containment performance below 0.05 ppm SF6 at recommended face velocity
No visible defects in powder coating or epoxy liner surface
All electrical connections tested for insulation resistance (> 1 M ohm)

Attachments

4 files
TDS-FH-001_Drawing_Rev2.pdf
2.4 MB • Jan 22, 2025
FH-001_Product_Photo.jpg
1.8 MB • Jan 22, 2025
FH-001_3D_Model.dwg
5.1 MB • Jan 20, 2025
ASHRAE_110_Test_Report.pdf
3.2 MB • Jan 18, 2025

Review History

2 revisions
Rev 2 — Approved
Technical Review by Suresh Iyer — Approved with no comments. Height updated from 2350mm to 2400mm, face velocity tolerance (±0.05 m/s) added per ASHRAE 110 requirements.
Jan 25, 2025  •  Reviewer: Suresh Iyer  •  Author: Ananya Patel
Rev 2 — Submitted for Review
Updated dimensions and added face velocity tolerance specification. Drawing Rev2 attached with corrected overall height.
Jan 22, 2025  •  Ananya Patel
Rev 1 — Approved
Safety Review by Dr. Ramesh Iyer — Approved with minor comment: "Recommend updating overall height to 2400mm for ergonomic sash operation."
Jan 20, 2025  •  Reviewer: Dr. Ramesh Iyer  •  Author: Ananya Patel
Rev 1 — Submitted for Review
Initial TDS creation with all technical specifications, standards compliance details, and product drawings attached.
Jan 18, 2025  •  Ananya Patel
TDS Created
Initial draft created for Fume Hood 4ft Standard Specifications under project PRJ-2025-001.
Jan 15, 2025  •  Ananya Patel

Related Items

5 linked items