Edit Reviewer Allocation In Progress
TDS-FH-024 : Fume Hood 6ft Walk-in Specifications — Assigned to Suresh Iyer
Current Allocation Details
Review Progress
Review Started
Feb 12, 2025
Days Elapsed
1 of 8 business days
Comments Added
2
Last Activity
Feb 12, 2025 at 14:30
Checklist Progress
6 of 10 items completed
Recent Review Comments
SI
Suresh Iyer Feb 12, 14:30
Air face velocity specification looks good at 0.5 m/s. Need to verify containment test results against ASHRAE 110.
SI
Suresh Iyer Feb 12, 11:15
Material specification for SS 304 liner is consistent with IS 4209 requirements. Epoxy coating thickness needs clarification.
Reviewer Reassignment
Transfer Existing Comments
Carry over all 2 review comments to the new reviewer
Transfer Checklist Progress
Carry over the 60% checklist completion to the new reviewer
Notify Original Reviewer
Send notification to Suresh Iyer about the reassignment
Deadline Extension
Extensions beyond 2 business days require approval
No approval required (within 2-day limit)
Approval from Engineering Manager needed if extension exceeds 2 business days
Audit Trail
Review started by Suresh Iyer
Allocation created and assigned to Suresh Iyer
TDS-FH-024 submitted for review